Skip navigation
Please use this identifier to cite or link to this item: https://libeldoc.bsuir.by/handle/123456789/48920
Full metadata record
DC FieldValueLanguage
dc.contributor.authorБагаев, С. И.-
dc.contributor.authorСмягликов, И. П.-
dc.contributor.authorЛазарук, С. К.-
dc.contributor.authorКупреева, О. В.-
dc.coverage.spatialМинск-
dc.date.accessioned2022-11-08T08:57:13Z-
dc.date.available2022-11-08T08:57:13Z-
dc.date.issued2021-
dc.identifier.citationБагаев, С. И. Биоактивные покрытия на основе оксидов титана и тантала для титановых имплантатов=Bioactive coatings based on titanium and tantalum oxides for titanium implants / С. И. Багаев [и др.] // Современные методы и технологии создания и обработки материалов : сборник научных трудов : в 2 кн. / Физико-технический институт Национальной академии наук Беларуси. – Минск : ФТИ НАН Беларуси, 2021. – Кн. 1. – С. 250-263.ru_RU
dc.identifier.urihttps://libeldoc.bsuir.by/handle/123456789/48920-
dc.description.abstractПриведены результаты экспериментальных исследований морфологии и свойств биоактивных покрытий, формируемых электрохимическими способами на титановых имплантатах, в виде электретных покрытий на основе оксидов тантала и оксидных покрытий, с высоким содержанием кальция и фосфора. Показано, что электретный заряд в покрытии сохраняется на протяжении не менее 90 дней и характеризуется средней плотностью заряда 4,4•10–4 Кл/м2. Кальций-фосфатные покрытия характеризуются поверхностной пористостью 5–15 %, размерами пор – 0,3–9,4 мкм, отношением концентраций Ca/P – 0,7–2,0, шероховатостью поверхности Ra до 1,7–1,9 мкм в зависимости от режимов плазменно-электролитической обработки. Введение в состав оксидного покрытия танталовых соединений снижает коррозионный ток более чем на порядок.ru_RU
dc.language.isoruru_RU
dc.publisherФТИ НАН Беларусиru_RU
dc.subjectпубликации ученыхru_RU
dc.subjectимплантатыru_RU
dc.subjectкальций-фосфатное покрытиеru_RU
dc.subjectбиоактивные свойстваru_RU
dc.subjectimplantru_RU
dc.subjectcalcium phosphate coatingru_RU
dc.subjectbioactive propertiesru_RU
dc.titleБиоактивные покрытия на основе оксидов титана и тантала для титановых имплантатовru_RU
dc.title.alternativeBioactive coatings based on titanium and tantalum oxides for titanium implantsru_RU
dc.typeArticleru_RU
local.description.annotationThe results of experimental studies of the morphology and properties of bioactive coatings formed by electrochemical methods on titanium implants, in the form of electret coatings based on tantalum oxides and oxide coatings with a high content of calcium and phosphorus, are presented. It is shown that the electret charge in the coating is maintained for at least 90 days and is characterized by an average charge density of 4.4·10 –4 C/m 2 . Calcium phosphate coatings are characterized by a surface porosity of 5–15 %, pore size – 0.3– 9.4 µm, Ca/P concentration ratio –0.7–2.0, surface roughness Ra up to 1.7–1.9 µm, depending on conditions of plasma electrolytic treatment. The introduction of tantalum compounds into the oxide coating reduces the corrosion current by more than an order of magnitude.ru_RU
local.description.annotationThe results of experimental studies of the morphology and properties of bioactive coatings formed by electrochemical methods on titanium implants, in the form of electret coatings based on tantalum oxides and oxide coatings with a high content of calcium and phosphorus, are presented. It is shown that the electret charge in the coating is maintained for at least 90 days and is characterized by an average charge density of 4.4·10 –4 C/m 2 . Calcium phosphate coatings are characterized by a surface porosity of 5–15 %, pore size – 0.3– 9.4 µm, Ca/P concentration ratio –0.7–2.0, surface roughness Ra up to 1.7–1.9 µm, depending on conditions of plasma electrolytic treatment. The introduction of tantalum compounds into the oxide coating reduces the corrosion current by more than an order of magnitude.-
Appears in Collections:Публикации в изданиях Республики Беларусь

Files in This Item:
File Description SizeFormat 
Bagayev_Bioaktivnyye.pdf5.77 MBAdobe PDFView/Open
Show simple item record Google Scholar

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.